Spectral Measurements

Introduction

Light is an electromagnetic wave. The wave nature of light was revealed by the combined work of Poisson, Fresnel, and Maxwell during the 19th century.  Maxwell developed a set of equations that describe the connection between the electric and magnetic fields and how light waves propagate in space. 

The two fundamental parameters of waves are the frequency (or temporal period) and the spatial  wavelength. The frequency describes how rapidly each point is moving up and down. High frequency means that there is a short time between each cycle and low frequency means that there is a long time between each cycle. The wavelength of a wave describes how far two adjacent peaks are. Long wavelength means the peaks are separated far from each other, and short wavelength means the peaks are closer. The relation between and the frequency of the wave and its wavevector is called the dispersion curve and depends on the material’s optical properties. 

One of the most important properties of waves is the ability to interfere with other waves leading to constructive interference when two waves are in phase, so they are enhancing each other, or destructive interference when the two waves are out of phase resulting in a weaker signal than the waves on their own. 

A periodic series of sources generate constructive beams at specific angles. This happens when the distance differences between each adjacent source are an integer number of the wavelength. The phenomenon, which is illustrated in Fig. 1(a), is known Bragg equation that reads:

\sin(\theta_{in})+\sin(\theta_{out})=\frac{n \lambda}{\Lambda}

Where \lambda is the wavelength, \Lambda is the distance between the sources, \mathrm n is the order number, and \theta_n is the angle for each diffracted order. The different diffraction orders lead to several beams where the zero order is directed forward and the other orders have larger angles. Instead of a series of sources, it is possible to illuminate a grating of slits with a plane wave. This is illustrated in Fig. 1(b), where the different orders direct the beam away from its original direction. Note that when increasing the index of refraction enough, the value of the right-hand side will increase above unity. In that case, no light will pass to the other side.

The diffraction efficiencies of each order can be calculated by the Fourier transform of the shape of the groove in the grating. If the diffraction grating is composed of square slits, then the efficiency of the n diffraction order is:

\eta_n=\frac{4}{\pi^2 n^2}(\pi n D)

Where D is the filling factor of the grating. Namely, D is the ratio between how much light is not blocked compared to how much light hits the grating. If the space between adjacent slits is twice the width of the slit, we will get D=50%.

When the grating is composed of simple slits, a similar number of positive and negative orders  will appear. It is possible to design a reflected blazed grating illustrated in Fig. 2. This grating has a preferable side leading to a single diffraction grating.

With this type of grating, it is possible to increase the efficiency of the first order beyond 80%. Such grating has high diffraction efficiency where the diffraction angle is a function of the wavelength. Therefore, different spectra components are diffracted in different directions. By placing a lens at the focal distance from the grating, the different angles are mapped to different positions at the Fourier plane. As illustrated in Fig. 3., placing a camera or detector array at the Fourier plane enables us to measure the spectrum of the input signal.

Spectrometers

The spectrometer is a device for measuring the spectrum of an input beam. It is based on separating the input beams into their different spectral components in space and measuring their intensity with an array of detectors. The spectral separation is usually done by resorting to either a prism or a grating. The prism or the grating gives a different angle to each spectral component and a lens at the focal length maps these angles into different locations on the detector array. The dispersion of a prism and a grating is shown in Fig. 1.
Fig. 1. Dispersion of a prism (left) and a grating (right), which is the basis for any spectrometer.

The spectral range of the spectrometer defines the bandwidth that the spectrometer can measure. The resolution of the spectrometer defines the minimal spectral feature that the spectrometer can resolve. Usually, there is a trade-off between the spectral range of the spectrometer and its spectral resolution. We will focus here on two types of spectrometers, one is grating-based and the other is prism based. We will also discuss some less common spectrometers, such as Fabry–Pérot` spectrometers and their properties. 

Diffraction grating-based spectrometer

As we discussed earlier, in diffraction gratings, the reflected angle, \theta , from the grating depends on the wavelength according to the diffraction Bragg equation:

\sin(\theta_{in})+\sin(\theta_{out})=\frac{n \lambda}{\Lambda}

Where \Lambda is the grating separation, \lambda is the wavelength and n is the reflected order. In interferometers, the first constructive order, where \mathrm n=1 is used for increased efficiency. After the grating, we place a lens at its focal distance \mathrm f, and after another focal distance, we place an array of detectors. Thus, the lens maps the different angles into different locations on the detector array. We demonstrate this spectrometer in Widget 1, where the user can build a simple spectrometer and then change the input wavelength and observe how the output changes.

Design considerations of spectrometers:

When designing a spectrometer, several considerations must be addressed. First is the spectral range of the spectrometer. The spectral range is defined by the spectral response of the detectors and by the spectrometer geometry. The spectral response of the detectors, or other components, can limit the spectrometer bandwidth. Therefore, some spectrometers are built with different types of detectors to increase the spectral range of the spectrometer.  The geometry of the spectrometer set which wavelength will reach the edges of the detector array, so any wavelength outside of this range will not reach the detectors. The size of the detectors, denoted as d in Fig. 2 leads to the spectral bandwidth \Delta \lambda of:

\Delta \lambda=\frac{d}{\Lambda f}

When we assume small angles and a zero incident angle:

\sin(\theta+\Delta \theta)=\sin(\theta)+\Delta \theta \cos(\theta)

To increase the bandwidth, the size of the array of detectors, d, can be increased so we will be able to measure spectral components that missed the smaller detector. We can also reduce the focal length, f, and place the lens closer to the grating so a wider range of spectral components will reach the detector. Lastly, we can resortthe grating with shorter periodicity. Also, the lens aperture can limit the spectral range of the spectrometer if the detector array is long enough, some spectral components can miss the lens and will not reach the detector. However, since it is cheaper and easier to find bigger lenses, it is usually not the limiting factor. A common technique to increase the spectral range is to rotate the grating during the measurement which increases the effective detector arrays but leads to a slow measurement. It is also possible to set the angle according to the wavelength we want to measure.

Second, we consider the spectral resolution of the spectrometer. The spectral resolution is set by the number of grooves in the grating that the light interacts with and by the spectrometer geometry. The diffraction Bragg equation, Eq. 1, is accurate when the light interacts with a large number of grooves of the grating. Then, a single spectral component will result in a narrow output beam which will interact with a single detector in the array. However, when the light interacts with a small number of lines in the grating, each spectral component results in a wide beam leading to a smaller spectral resolution. The figure-of-merit here is that to obtain a δλ resolution, and the light must interact with at least δλ/λ lines of the grating. The wavelength sets the grating periodicity we need to measure, and usually, it is not easy to change. In addition, the size of the input beam is also set by the system. Therefore, there are two ways to increase the number of lines which the light interacts with. The first way is to resort to a beam expander similar to a telescope. However, this increase in the complexity of the system increases its size, which is already a problem in sensitive spectrometers, and it increases the sensitivity to any input misalignments. The second way to increase the number of lines is to direct the light to reach the grating at a shallow angle and make the grating as large as possible. Many spectrometers resort to both ways to increase the resolution to its maximum.

Fig. 2. Schematics of a simple spectrometer based on a grating.

The spectrometer geometry also has a strong influence on the spectral resolution of the spectrometer. Even if two components lead to two narrow beams which do not overlap, they still need to reach different detectors in the array for them to be separately resolved. Thus, if the separation between two detectors is x the resolution of the spectrometer δλ is set according to:

\delta \lambda=\frac{\Lambda f}{x}

Thus, to increase the resolution, we should use smaller detectors, lenses with longer focal lengths, or grating with fewer lines per mm. These requirements are the opposite of the resolution requirements and therefore, the spectral range of the spectrometer and its spectral resolution has opposite design considerations. Usually, to increase the spectral resolution, spectrometers usually use big large lenses with a focal length of a few meters, reducing their spectral range. To set an entire spectrometer in a small box, the engineers are resorting to mirrors for folding a long distance in a small footprint.

Third, when the angle of the reflected light from a grating is different from the incident angle, the beam shape is distorted. For example, if the input beam has a circular cross-section shape, then the cross-section of the reflected beam will become an ellipse shape. This may lead to aberrations in the lens focusing and distortion in the output spectra. These aberrations are easy to compensate for, but another option is to hit the grating so the beam will be reflected close to the incident beam. This angle is called the Littrow angle and is found according to:

\sin(\theta_L)=\frac{\lambda n}{2 \Lambda}

This is obtained by resorting to the complete Bragg condition for an arbitrary incident angle as:

\sin(\theta_{in})+\sin(\theta_{out})=\frac{n \lambda}{\Lambda}

And setting \theta_{in}=\theta_{out}. Thus the shape of the input beam is preserved after the reflection from the grating. The Litro condition sets the input angle so it is not a free parameter anymore and therefore, in order to increase the resolution, a denser grating must be used.

Prism-based spectrometer

Another way to design a spectrometer is by a prism. However, the dispersion of a prism is not as high as a diffraction grating, and therefore, it has a lower spectral resolution but an increased spectral bandwidth. The dispersion of a 60 degrees prism of fused silica glass is a function of the wavelength, but it can be as high as 0.03 per nm. Therefore, for the same focal length and the same detectors, the spectrometer will have a resolution that is two orders of magnitude less. The benefit of prism-based spectrometers is the increased efficiency since such a spectrometer can have close to 99% efficiency. Much higher than the grating, which can be as low as 50%. 

The dispersion of the prism as a function of the prism size and prism geometry is:

\sin\left(a_{out}\right)\;=\;n_{2\left(\lambda\right)}\sin\left(\frac{{}^atop}2\right)

Where n_{2\left(\lambda\right)} is the index of refraction of the material, which is a function of the input wavelength due to the dispersion, and \alpha_{top} is the angle at the top of the prism. As shown from this equation, by increasing the angle of the prism, we can increase the resolution. However, we note that we cannot increase the angle too much due to total internal reflection. When the right-hand side of this equation will increase beyond unity, there will be no answer for this equation. In this case, we will have total internal reflection which will prevent the rays from exiting the prism. The size of the prism is important for the resolution since it allows us to resort to larger lenses for better spectral bandwidth and higher resolution. The schematic of the spectrometer is shown in Fig. 3. The resulting measurements of such a spectrometer are shown in Fig. 4.

The other benefits of such spectrometers are the increased spectral bandwidth and the robustness of the system. Therefore, such spectrometers are suitable for analyzing weak signals over a wider bandwidth, where the resolution is less important. It is important to note that such spectrometers are not based on reflection from a grating but on transmission through the prism, and therefore, from an engineering point of view, they are more challenging to fold a large optical scheme into a small box.

Fig. 3. Schematics of basic spectrometer based on prism measuring five laser beams with different wavelengths.
Fig. 4. Simulated measurements of the output from the spectrometer showing the five different spots corresponding to the five different laser beams.

Exercise 1:

The student will measure the resolution of two spectrometers: grating-based and prism-based spectrometers.

  • Load the file “spectrometer_ex1.opt”. In the file, there are two spectrometers, one based on a grating and the other on a prism.
  • Run the simulation and measure the output of the two spectrometers.
  • Change the input wavelength of both spectrometers and check when there is a notable change in the output, namely, for what wavelength shift the light reaches a different pixel in the detector.
  • Find the spectral bandwidth of the spectrometer by checking what is the longest and shortest wavelengths that can still enter the detector in both spectrometers.
  • Change the focal length of the lens to 50 mm and reposition it accordingly. What is the spectral resolution and the spectral bandwidth of both spectrometers?

Monochromator

Monochromators are devices similar to a spectrometer that is designed to filter a specific spectral component out of an input. Some of them have tunability in the output wavelength and bandwidth, while others are fixed. In these devices, the detector array is replaced with a slit, so only a narrow wavelength can pass and propagate toward the output. Therefore, monochromators have optical input that are commonly broadband and a narrowband optical output. By changing the location of the slit, the user can tune the output wavelength and by changing the width of the slit, the user can set the bandwidth of the output. Since spectrometers and monochromators are similar, many devices have both, i.e., a spectrometer which is also a monochromator, and vice-versa. A schematic of a monochromator that is based on a grating is shown in Fig. 5. It is also possible to design a monochromator based on prisms. Such monochromators has wider bandwidth but with lower spectral resolution.

Fig. 5. Schematics of a monochromator showing how to isolate a single wavelength out of broadband input signal.

Pulse shaper

The monochromator is a basic spectral manipulator that blocks all the spectral components other than  a single spectral component that is directed as the output. Pulse shapers are more advanced than monochromators, and with them, the user can impose a complex function of both phase and intensity on the spectrum of the input signal. To obtain a pulse shaper, the slit is replaced with a spatial light modulator (SLM) that can impose an arbitrary phase or intensity as a function of space. 

Pulse shapers have the same resolution and bandwidth considerations as a spectrometer, but now it is impossible to scan the spectral range slowly. Pulse shapers are designed to alter the spectrum of a pulse and thus, shape the pulse. Therefore, we need to address the entire spectral bandwidth in a single interaction.

Usually, a pulse shaper needs lower resolution but broader spectral bandwidth than spectrometers. These setups are possible only when the distances between the elements are short since the system does not compensate for diffraction. The benefits of this system are that it is easier to align and it is possible to tune the system bandwidth and resolution only by changing the distances between the elements and without the need to replace the lenses each time. Schematics of a pulse shaper is shown in Fig. 6, where the different wavelengths are separated in space and then combined again into a single beam.

Fig. 6. Simulations and ray scheme of a pulse shaper. This scheme disperses the wavelengths in space and then combines them again. By setting different phases and intensities in the spectrum, we design different pulses at the output.

Exercise 2:

Pulse shaper and monochromator

  • Load the file “spectrometer_ex2.opt” and run the simulation.
  • Measure the resolution of the pulse shaper.
  • Replace the detector with a slit and measure the spectral bandwidth of the pulse shaper.
  • What is the narrowest output in this monochromator?
  • Load the file “spectrometer_ex3.opt” , run the simulation and repeat bullets 2-4.